Trouble Shooting; Note Of Care - Wohler TI 410 Operating Manual

Dew-point indicator
Hide thumbs Also See for TI 410:
Table of Contents

Advertisement

Available languages

Available languages

8
Trouble Shoot
The following table shows possible
Indicator
Battery symbol appears in
the display.
LOW
HIGH
ERR
9

Note of care

ing
fa
ilures and explains how to repair them.
Possible reason
The necessary battery volt-
age is undercut.
T < - 40°C
T > 100°C
Measuring error during the
stabilization phase
The sensitivity of the TI 410 depends very strongly
on the condition of the sensor plate. If the sensor
plate is dirty, the sensitivity is lowered strongly.
Use a cloth dampened with aceton for cleaning.
The cleaning must be carefully done so that the
temperature sensor is not damaged. Hand contac
with the sensor plate should
m
uch as possible.

Trouble Shooting

Trouble shooting
Change Battery.
Temperature is out of
range.
Temperature is out of
range.
Switch off the device.
Wait until the sensor is dry.
Switch on the device.
be minimized as
EN
t
23