Caution
Equipment Damage — Before connecting the
ergometer to the power line, make sure that the
voltage rating of your local power line matches the
value indicated on the device nameplate. The
nameplate is located on the back of the ergometer,
at the bottom.
g
f
e
g
PORT 2
f
PORT 1
NOT CONNECTED
e
Figure 2-5 Connector panel
a
Power input, instrument fuses
b
Digital interface RS232
c
Analog I/O
d
For Service
e
Not used
f
Digital interface RS232
g
Digital interface RS232 (for Service only)
2018112-001 Revision B
Setup and Mains Connection
a
b
c
d
b
8
7
6
c
3
1
3
1
5
4
5
4
2
2
PORT 3
ANALOG
d
SERVICE
eBike with Control Terminal PC
Mains Connection
Stand in front of the ergometer and grasp both handles
of the handlebar.
Tip the ergometer forward all the way and carefully
lay it on its handlebar.
Now you can access the connector panel on the underside
of the ergometer.
Connect the power cord to socket a (Figure 2-5).
Secure the connector with the supplied lock a.
Install the strain relief as shown in Figure 2-5. Please
note that different strain reliefs are supplied for
various cable sizes.
Connection to Electrocardiographs
The software supports communication with the following
electrocardiographs:
electrocardiographs (digital control):
electro-
connection
cardiograph
cable
CardioSys
22336203
CardioSoft
22336203
CASE
22336203
CASE 8000
22336203
MicroLab
22336203
CardioSmart ST*) 22338002
MAC 1200 ST
22338002
*) eBike with integrated BP module only
electrocardiographs (analog control):
electro-
connection
cardiograph
cable
MAC 5000 ST
2008110-001
2008114-001
Upon delivery, the eBike is configured for
communication with digitally controlled
electrocardiographs.
For use with electrocardiographs with an analog control
system (MAC 5000 ST), the following settings are
required (see section 6 "Configuration"):
– at Default Mode: PC Operation
– at EKG-Typ: Analog / Digital
Install a strain relief to secure the cable connection.
connector on
eBike
port 1, f
port 1, f
port 1, f
port 1, f
port 1, f
port 3, b
port 3, b
connector on
eBike
ANALOG, c