EKG-Typ
0 0 3
S e l e c t
Figure 6-13 Entering the password
EKG-Typ
A n a l o g / D i g i t a l
D i g i t a l
S e l e c t
Figure 6-14 Changeover from digital to analog control
R o t a t i o n
M i n .
0 . . . 7 0
M a x .
5 0 . . . 1 3 0
S e l e c t
Figure 6-15 Adjusting RPM limit values
2018112-001 Revision B
Configuration
5 4
/ m i n
6 5
/ m i n
eBike with Control Terminal PC
EKG-Typ
In this menu you toggle between digital and analog
control of the ergometer. Access to this menu is protected
with a password.
Using the arrow buttons, enter the value "003" in the
password box and confirm with Select.
Now you will see Figure 6-14.
Using the arrow buttons, choose the "EKG-Typ", then
confirm with Select.
Rotation
In this menu you choose the limit values for an "RPM
alarm", when the LED for high or low speed (RPM)
lights up on the speed display.
Use the arrow buttons to choose the minimum or
maximum RPM parameter and confirm with Select.
Change the RPM limit value with the arrow buttons
and confirm with Select.
The minimum value can be adjusted between 0 and 70
RPM, the maximum value between 50 and 130. With the
limit values shown in Figure 6-15, the LED for the low
speed lights up if the patient pedals below 54 RPM, the
LED for the high speed lights up above 65 RPM.
Pulse Indication
With the menu item "Display pulse" you enable and
disable indication of the pulse rate.
25