GE eBike with Control Terminal PC Operator's Manual page 53

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 39
P r o g r a m s
1 .
W H O
2 .
B A L
3 .
H o l l m a n n
4 .
S T D . F r a n c e
5 .
S t a n d a r d
S e l e c t
Figure 4-4 Protocol menu
a
b
2 5
W a t t
m i n
/
m m H g
S t a r t
Figure 4-5 Ergometry mode
a
Blood-pressure readings (systolic, diastolic)
b
Initial load (appears 15 s after activation of Start)
c
Duration of exercise test
d
Pulse rate (configurable)
e
Speed (revs/min)
2018112-001 Revision B
Running an Exercise Test
c
d
e
0
0
/ m i n
/ m i n
eBike with Control Terminal PC
4.3 Exercise Test
Selecting and Starting the Operating Mode
In this operating mode, the protocols stored in the
ergometer will control the device.
Using the right and left softkey in the main menu,
position the bar cursor on Ergometry, then press the
softkey in the middle to Select the option.
Note
The device may be configured to default to one of
the operating modes. In this case you will see the
initial screen of the selected operating mode instead
of the main menu. You can display the main menu
with the
button (see section 6
"Configuration").
The protocol menu appears (Figure 4-4). The bar cursor
highlights the test protocol used last. With the right and
left softkeys you can choose among the 15 protocols.
With the softkey in the middle you Select the option and
display the initial screen (Figure 4-5).
When you have finished preparing the patient, Start
the test with the softkey in the middle.
The ergometer will emit three beeps and the speed
display will blink three times to indicate the beginning of
the test. The exercise test clock (c, Figure 4-5) will start
at the same time. When, after approx. 15 s, the initial load
appears at b, the patient should begin pedaling.
Note
The protocol details are listed in the Appendix.
Refer to section "Configuration" on page 21 for
protocol editing instructions.
17

Advertisement

Table of Contents
loading

Table of Contents