Dcu 220 Replacement; Dispatch Of The Defective Dcu 220; Commissioning Of Dcu 220 Rental Device; Commissioning Of Repaired Dcu 220 - Bosch DCU 220 Repair Instructions

Diagnostic control unit
Hide thumbs Also See for DCU 220:
Table of Contents

Advertisement

Available languages

Available languages

7.

DCU 220 replacement

Availability
Processing via the Service Center Repair in Munich
functions for Europe and the Middle East.
Processing time
The processing time from acceptance of incoming
goods at the Service Center Repair until return ship-
ment is roughly 10 working days.
7.1

Dispatch of the defective DCU 220

i
DCU 220 involves delivery without storage batteries,
hard disk, stylus, power supply unit and accessories.
!
First check if a hardware fault or software fault is
present. DCU 220must only be sent in after a hard-
ware fault.
1. Remove both storage batteries from the defective
DCU 220.
2. Remove the hard disk from the defective DCU 220.
3. Remove the stylus from the defective DCU 220.
4. Return defective DCU 220 via Coremannet (for re-
placement purposes) or Bosch Diagnostics (liability
for defective goods) in the original packaging or in
equivalent packaging.
i
Submission address and form available on EDIS.
7.2
Commissioning of DCU 220 rental
device
i
The DCU 220 rental device is supplied without stor-
age batteries, hard disk, stylus, power supply unit
and accessories.
1. Install both storage batteries on the DCU 220 stor-
age equipment.
2. Install the hard disk in the DCU 220 rental device.
3. Install the stylus in the DCU 220 rental device.
4. Switch on DCU 220 rental device and installations
and licenses.
Robert Bosch GmbH
7.3

Commissioning of repaired DCU 220

1. Remove both storage batteries on the DCU 220
rental device.
2. Remove the hard disk from the DCU 220 rental
device.
3. Remove the stylus from the DCU 220 rental device.
4. Install both storage batteries on the repaired
DCU 220 rental device.
5. Install the hard disk in the repaired DCU 220.
6. Install the stylus on the repaired DCU 220.
7. Switch on the repaired DCU 220 installations and
check licenses.
7.4
Comment concerning material defect
liability
!
Before sending in the DCU 220 in respect of lia-
bility for defects the test steps on the checklist
1 689 980 335
must be executed. Material defect li-
ability claims will be rejected if it is established that
the necessary steps have not been undertaken.
The completed Checklist must always be included
with the material defect liability claim (in paper form
returned with the product, and as an attachment in
BWS online). The Checklist
as a form in EDIS and can be filled out on a PC/laptop
before saving as a PDF document and printed out.
In the case of justified material defect liability claims,
the time employed for working through the Checklist
will be reimbursed.
Measure
Working through the Checklist.
Working through Checklist and perform-
ing operating system recovery.
| DCU 220 | 31
en
1 689 980 335
is available
Working time
max. 0.15 h
max. 0.50 h
|
1 689 975 264
2015-06-02

Hide quick links:

Advertisement

Chapters

Table of Contents
loading

Table of Contents