Turbo-V 1K-G Switching off

To switch off the Turbo-V 1K-G it is necessary to remove the mains.
The integrated controller immediately stops the pump.
WARNING!
The Turbo-V controller must be powered with 3-wire power cord (see orderable
parts table) and plug (internationally approved) for user's safety. Use this
power cord and plug in conjunction with a properly grounded power socket to
avoid electrical shock and to satisfy CE requirements. High voltage developed
in the controller can cause severe injury or death. Before servicing the unit,
disconnect the input power cable.

Emergency Stop

To immediately stop the Turbo-V 1K-G in an emergency condition it
is necessary to remove the supply cable from the mains plug.