Generator Battery Issues

LOW BATTERY/END OF SERVICE INDICATIONS IN OR
If a generator displays a low battery or End of Service (EOS) indicator while
implanting, the possible causes include the following:
Prior to Surgery:
• Generator has been recently
exposed to low storage
temperatures
• Defective generator
Wait approximately
30 minutes with generator
at room temperature
Error Message
• "Generator Battery N EOS..."
• "Generator Battery EOS..."
• "Frequent monitoring recommended.
The Intensified Followup Indicator (IFI) is set..."
• "Generator disabled..."
• Perform System Diagnostics or Generator
Diagnostics with device in sterile package
• Record results
Same battery status
indicator or warning
message as above
1
During Surgery:
• Electrosurgical equipment used near
the generator
• Generator exposed to electrostatic
discharge (ESD)
Diagnostic
Results
2
Number of
attempts
• Lead Impedance "any"
• Impedance Value "any"
• IFI = No
Proceed with implant
Contact LivaNova
Technical Support
77