In order for the generator to accurately detect heartbeats, the heartbeat
detection must be set for the individual patient. Manually select from a range
of heartbeat detection sensitivity values: 1 (least sensitive; for use with largest
amplitude ECG signals) to 5 (most sensitive; for use with smallest amplitude
ECG signals). The setting will not change unless manually programmed to a
different value.
When Detection is enabled, the software will walk you through heartbeat
detection setting verification and AutoStim Threshold selection.
To Verify Heartbeat Detection:
1. Select
VERIFY
rate detected by the generator or to change the Heartbeat Detection setting
2. Place the Wand over the generator and press
Keep the Wand over the generator during the entire Verify Heartbeat
Detection process. The generator will transmit a signal and the Programmer
will display the detected heart rate in beats per minute (BPM) for up to
two minutes
on the Detection Tab to confirm the accuracy of the heart
START
to begin the test.