Select AutoStim Threshold
The AutoStim Threshold is the percent heart rate increase over baseline that is
required to trigger automatic stimulation (AutoStim). Values range from 20%
(most sensitive) to 70% (least sensitive) and should refl ect the magnitude of
heart rate increase associated with the patient's seizure.
To select the AutoStim Threshold:
1. From the Summary Screen, tap on the
OR tap on the
of the screen
2. On the Parameters Screen, tap on the Detection Tab and confi rm that
Tachycardia Detection is enabled
3. Select an AutoStim Threshold. The threshold is the amount of heart rate
increase needed to trigger an AutoStim (in the range of 20% to
70% increase from baseline)
4. Tap
APPLY CHANGES
PARAMETERS