Wand (M2000)
The wand is designed to communicate wirelessly with the programmer
(M3000) at distances up to 3 meters (10 feet) under most conditions.
If communication is unstable, use the supplied USB cable to connect the
wand and the programmer instead.
Hold the wand no more than 1 inch away from the generator in order to
maintain good communication.
The wand operates on two AA lithium or alkaline batteries.
Magnet
The VNS Therapy® System has a patient controlled feature that can be activated
with a specially designed magnet.
The magnet can be used to:
• Attempt to reduce the intensity of or abort an oncoming seizure or
seizure in progress
• Temporarily inhibit stimulation
• Verify the generator is providing stimulation
• Reset the generator (in combination with the programmer and wand)
Programmer
Programming
Programmer
Charger
USB Cable
The magnet is available in two styles.
Pager Style
Magnet
Watch Style
Magnet
Wand
Wand