NOTE:
Check box only appears if the programmer does not have a preferred
wand. If a preferred wand is selected, the programmer will search and
automatically connect to the preferred wand.
After a successful interrogation, the Summary Screen will display.
From this screen, you can perform the following functions:
•
View generator ID information, including model number and serial number
•
View and edit patient data, such as patient ID and implant date
•
View last known diagnostics data, such as lead impedance and battery status
•
Change settings to generator parameters, such as Normal, Magnet, AutoStim,
or Detection settings
•
Perform Diagnostics
•
View events and trends such as magnet activations and daily average
AutoStims
•
Access device history, including parameter settings associated with prior
offi ce visits
•
Interrogate the generator again to verify parameters or refresh data
•
End programming session
•
Access other software options