Electrical installation

Model
Market
Europe
Stillbath
North America
Japan
Europe
Hydrosound
North America
Japan
Europe
Hydromassage
North America
Japan
First check that the electrical supply
corresponds with the information on
the label on the backside of the
panel.
The unit must be permanently
installed.
The unit must be continuously
powered and connected to separate
fuse with ground fault circuit
interrupter.
Main switch must be placed close to
the tub and visable for the user.
Power supply is activated via the
main switch.
Fuses according to EN60898.

DIP Switches for Autofill

The model is equipped with dip switches,
which are located on the PCB and can be
adjusted as in the tables below:
PCB located on the backside of the panel
front.
1
OFF
Not used
2
ON
Auto-fill activated
2
OFF
Auto-fill deactivated
3
OFF
0 seconds*
4
OFF
3
ON
60 seconds*
4
OFF
3
OFF
120 seconds*
4
ON
3
ON
180 seconds*
4
ON
*the amount of seconds of extended filling time
from that the 0-level is fulfilled.
6
Voltage
230/240V single phase AC
120V single phase AC
100V single phase AC
230/240V single phase AC
120V single phase AC
100V single phase AC
230/240V single phase AC
120V single phase AC
100V single phase AC
The unit shall be electrically connected
with the following equipment
(not included in delivery):
Separate fuse.
Ground fault circuit
interrupter (10mA)
Safety switch, according to
regulations and local codes.
Mains cable
(3x16AWG / 3x1.5 mm
All electrical installations must comply
with the US National Electrical Code and
the Canadian Electrical Code respectively.

DIP Switch for Cleaning

6
ON
Automatic hydroclean
Time set for effect of
cleaning
5
ON
10 minutes
5
OFF
2 minutes
6
OFF
No hydroclean

DIP Switch Hydro/No Hydro

7
ON
Hydrosound/
Hydromassage
7
OFF
Stillbath
Fuse
Frequency
(slowblow)
50Hz
6A
60Hz
10 A
50/60 Hz
10 A
50Hz
6 A
60Hz
10 A
50/60 Hz
10 A
50Hz
6 A
60Hz
10 A
50/60 Hz
10 A
Connect the wires as described below:
EUROPE
Neutral to the blue wire (N)
Phase to the brown wire (L)
Protective earth to
the yellow/green wire (PE)
USA
Neutral to the white wire (NEUT)
Phase to the black wire (HOT)
2
)
Protective earth to
the green wire (GND)
Power
consumption
max.700 VA
max. 700 VA
max. 700 VA