Honeywell Sensepoint XCD RFD Technical Handbook page 32

Remote flammable detector
Hide thumbs Also See for Sensepoint XCD RFD:
Table of Contents

Advertisement

Sensepoint XCD RFD Technical Manual
19.Promptly switch off the calibration span gas and remove the Sensepoint XCD
Gassing Cap from the sensor to allow the gas to disperse.
20.When the reading falls below 50% of the calibration gas level the display indicates a
countdown .
21. When the countdown is finished, the calibration procedure is complete.
22. The instrument returns to the 'Set CAL' menu. Activate the '
select another menu or select 'QuIT' to return to normal monitoring mode.
Note: Remember to always replace the Weather Protection and other accessories.