WARNING
Fuse replacement should be performed only by a qualified service personnel. See Chapter 4.
MAINTENANCE, Fuse.
Access to terminal and cable entry
l
The power supply terminals are behind the shield cover.
Guide the power cables through the gland. The terminals will accept wires of 2.5 mm
Always use cable finishings. For details, refer to the User's Manual IM 12A01F01-02EN.
Power supply terminals
l
Connect power supply to the power supply terminals.
For details, refer to the User's Manual IM 12A01F01-02EN.
Grounding the converter
l
For the safety of the user and to protect the instrument against interference, the housing must
always be connected to ground. This has to be done by a cable with large cross-sectional area.
This cable can be fixed to the rear of the housing or the internal ground connections. For details,
refer to the User's Manual IM 12A01F01-02EN.
WARNING
The minimum cross sectional area of the protective grounding wire should be 0.75 mm
CSA safety standard (Type: - AB), use cables with a cross section of 0.75 - 2.1 mm
<Start-up and Safety Precautions>
(M4 screw)
AC power terminals
(AC Version)
DC power terminals
(DC Version)
(M4 screw)
2
.
IM 12A01F02-01EN
2nd Edition : Apr. 19, 2019-00
13
2
(14 AWG).
2
. For