Sensing head diaphragm replacement
Hide thumbs Also See for Milltronics ILE-37:

Advertisement

Step Four: Remove LVDT Conditioner Card
1.
Use 2.5 mm hex screwdriver or Allen key to remove
the two screws, using your free hand to steady the
LVDT conditioner card.
CAUTION: Do not touch circuitry on LVDT conditioner card!
Step Five: Separate Inner Diaphragm Ring from ILE-37 Body
1.
Use 5 mm Allen key to remove the four screws connecting the inner diaphragm
ring to the ILE-37 body.
Milltronics ILE-37 Sensing Head Diaphragm Replacement
2.
Remove two nylon washers and two metal spacers.
3.
Remove LVDT conditioner card and place on a dry,
secure surface.
AG050425
APPLICATION GUIDE
2.
Separate the inner diaphragm ring
from the ILE-37 body.