Preventing damage to device

Handling measuring results

NOTICE!
Damage to device
► Ensure that no liquids enter the device. They can damage the
electronics.
► Switch off the device before disconnecting the power supply unit
from the mains socket.
► For devices with power supply operation: Disconnect the power
supply unit from the mains socket if you intend to not use the device
for a longer period of time. Only this way it can be ensured that the
device is de-energized.
► For devices with battery or rechargeable battery operation: If you do
not use the device for an extended period of time, remove batteries
or rechargeable batteries. Only this way it can be ensured that the
device is de-energized.
► Make sure not to drop the device.
► Do not expose the device to any impacts or vibrations.
► Perform function controls regularly as described in the relevant
section in this document. Do not operate the device if it is damaged
or not working properly.
► Ensure that there is no heat source in the immediate vicinity. Do not
expose to direct sunlight. The excessive temperature could damage
the electronics.
► Avoid rapid temperature fluctuations. When the device is
transported so that a temperature difference of more than 20 °C
occurs, it must stay turned off for at least 2 hours before it can be
turned on again. Otherwise, condensation water will form which can
damage the electronics.
► Use the device only in the intended ambient conditions.
► Store the device only in the intended storage conditions.
► Do not use aggressive or abrasive cleaning agents.
► Do not use organic solvents (e.g. white spirit or petroleum spirit).
CAUTION!
Patient hazard
In order to avoid misinterpretations, measuring results for medical use
must be displayed and used in SI units (weight: kilogrammes, length:
metres) only. Some devices offer the ability to display measuring
results in other units. This is only an additional function.
► Use the results exclusively in SI units.
► The use of measuring results in non-SI units is the sole responsibility
of the user.
NOTICE!
Inconsistent measuring results
► Before you electronically save measured values determined using
this device and use them further (e.g. in seca PC software or in an
EMR system), make sure that the measured values are plausible.
► If measured values are transmitted to seca PC software or an EMR
system, make sure prior to further use that the measured values are
plausible and are assigned to the correct patient.
English
Safety information • 7