Auto Calibration; Auto Calibration (For Aqt 200 Only) - Adam Equipment AQT Series Service Manual

Hide thumbs Also See for AQT Series:
Table of Contents

Advertisement

Cannot use Full Capacity
Not Linear
Off Center Loading error
Keyboard failure
Incomplete LCD display
Backlight does not turn on

10.0 AUTO CALIBRATION

• Calibration may be required when the balance is initially installed
or if the balance is moved from a location.
• With time and use, mechanical deviations may occur. Therefore
calibrate the balance whenever the best accuracy is required or
when any deviation is observed during weighing a test weight.
• Before calibrating it is necessary to have the proper calibration
weights. The weights should be OIML Class M1 type or better.
If you do not have proper weights do not attempt calibration.
Model #
AQT-200
Calibration
200g
Weight

10.1 AUTO CALIBRATION (FOR AQT 200 ONLY)

• Before entering calibration mode, press [On/Off] to turn the
power on for 1 minute.
Enter calibration mode
• Press [On/Off] to turn the power off.
• Press [Unit] first and then press [On/Off], while keeping the
[Unit] key pressed. The display will show the A/D value.
@Adam Equipment Company 2004
Over load Stops hitting pan support
or hitting bottom of load cell
Shipping screw not removed from base
Electronic problem on A/D
Load Cell Damaged
Overload stops hitting too soon
Load cell damaged
A/D damaged
Capacitor(s) on A/D PCB damaged
Overload Stops not correct
Load Cell damaged
Keys damaged
Cable to display/keyboard not connected
Poor contacts on LCD
Damaged LCD or circuit board
Backlight module damaged
Backlight power supply failed
AQT- 250
AQT-600
250g
600g
8
AQT-1500
AQT-2000
1500g
2000g
AQT-5000
5000g

Advertisement

Table of Contents
loading

This manual is also suitable for:

Aqt-200Aqt-250Aqt-1500Aqt-600Aqt-2000Aqt-5000