Cleaning & Maintenance; External Cleaning; Troubleshooting - Ametek Reichert 7 Series Installation Procedure

Table of Contents

Advertisement

Cleaning & Maintenance
Perform the following procedures as need to keep this instrument operating correctly.
CAUTION: DO NOT USE ALCOHOL, SOLVENTS OR STRONG CLEANING SOLUTIONS OR DAMAGE
TO THE UNIT WILL OCCUR.

External Cleaning

Remove power from the Reichert 7 and clean the external surfaces of the Chinrest using a clean, soft
cloth lightly moistened with a mild detergent solution (1 cc of liquid dish soap to one liter of clean, filtered
water (filtered below 5 microns)). Do not allow the mild detergent solution to drip inside the Reichert 7 or
Chinrest Assembly or damage may occur.
There is no scheduled maintenance required for this product.
Fuse Replacement
There are no fuses in this product.

Troubleshooting

The following chart provides details of common complaints and solutions for the Reichert 7 Series
Chinrest.
WARNING: DO NOT REMOVE THE OUTSIDE COVERS OF THE UNIT OR ATTEMPT TO REPAIR ANY
INTERNAL PARTS. REPAIR AND SERVICE OF THE UNIT MUST BE PERFORMED BY EXPERIENCED
REICHERT PERSONNEL OR AUTHORIZED DISTRIBUTORS WHO ARE TRAINED BY REICHERT.
Details
Chinrest does not
move Up or Down.
Chinrest moves left and
right and is not locked
in position.
16049-101 Rev. D
Probable Cause
• Chinrest is not connected to
the Reichert 7.
• Up/Down Switch is not
connected to Reichert 7.
• Circuit board in the Chinrest
assembly is defective.
• Chinrest mounting bracket is
loose.
Solution
• Attach connector to the Reichert 7 .
• Attach the connector from the switch
to the Reichert 7.
• Replace the Chinrest circuit board.
• Tighten mounting bracket on the bot-
tom of the Reichert 7.
9