Shenchen Pump
Click ' Calibration' button in main interface, enter flow rate calibration interface.
In dispensing mode, calibration interface data: dispensing volume and dispensing
time is the data set in main interface before. In transferring mode, dispensing
volume is current flow rate refer to dispensing time. Click ' Dispensing time' to
change the calibration volume.
Calibration process as below:
A. Confirm dispensing volume and time.
B. Click Start button to start the test, display countdown dispensing time, it will
stop automatically, and the numerical keyboard come out, you can input the actual
dispensing volume. After input the actual dispensing volume, it will ask whether
continue test ( suggest 3 times), choose 'YES', it will test again; choose 'NO', back
to the calibration interface.
C.
After several tests, actual dispensing volume display area display the average
volume, click 'Calibration' button, display calibrate successful.
D. If request higher accuracy, you can use the adjust function, click 'Add' or 'Dec'
button, micro adjust the dispensing volume, meet high accuracy dispensing.
Online Micro Adjust Dispensing Volume Process:
If the dispensing volume is not correct during working, this function can micro
adjust the volume online without affect the production line.
A) Click the Calibration button on the main interface, enter flow calibration
interface.
B) Now only the 'Add', 'Dec' and 'Esc' button can be used, other button is
Actual Vol.
0.0000ml
Disp. Vol.
10.00
ml
Test
Disp. Time
CAL
1.00
s
1.00
Cancel
Volume Adjustment
+0.0000ml
Add
Dec
Esc
- 8 -