DD+DIS148.05E
8

Electrical connections

Installation guidelines
IMPORTANT:
A ground fault interrupter (GFI) (I N = 30 mA) must be installed at a suitable point in
the circuit (in compliance with VDE 664).
Mains connection in the
installation room
Mains cable
(part of the delivery)
Network cable
(part of the delivery)
Main switch
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
12-2007
Electrical installations have to be done in accordance
with national regulations. Examples:
VDE
(Germany) Electrical installations in the
installation room must be in compliance with the
regulations IEC 364, VDE 0100, and VDE 0107.
UL
(US) Electrical installations in the installation
room must be in compliance with the regulations:
"National Electrical Code" (NEC) (NFPA70).
VDE
Double earthed pin outlet in compliance with
DIN 49441 and with CEE 7 standard cover V II.
UL
Earthed contact outlet for the NEMA 5-15 P plug
CEE7/7 - H05VV - F3G - 3 x 2.5 mm² with
Euro-
Version:
IEC C13 plug (female); cable length 3 m
US/CAN-
NEMA 5-15P - SJT, 3 x AWG 18 with
Version:
IEC C13 plug (female); Cable length 3 m
US
NEMA 5-15P
UTP; AWG24; Length: 5 m
Upon machine installation, it must be ensured that either
the mains connector or an all-pole circuit breaker for the
installation on site is located close to the machine and
easily accessible.
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
5366/100
Installation Planning
Europe
IEC C13
IEC C13
Figure 9
Agfa Company Confidential
1
2
3
CEE7/7
5366_chap11_001.cdr
Chapter 11 / 21