Effect Of Temperature On Working Pressure - Siemens SITRANS F M MAG 5100 W Operating Instructions Manual

Electromagnetic flowmeters
Hide thumbs Also See for SITRANS F M MAG 5100 W:
Table of Contents

Advertisement

Max. cable loop resistance
Cable glands on sensor
and transmitter
Special applications, e.g. low conductivity or electrical noise
Table 8- 7
Basic data
Ambient temperature
Cable parameter
8.3

Effect of temperature on working pressure

Effect of temperature on working pressure.
Table 8- 8
Flange
specifications
Sizes DN25 ... 2000
EN 1092-1
ANSI B16.5
AWWA C-207
SITRANS F M MAG 5100 W
Operating Instructions, 11/2010, A5E03063678-02
Media temperature:
< 100 °C
> 200 °C
M20x1.5 gland - Cable ø 5 ... 13 mm (0.20 ... 0.51 inches)
½ NPT gland - cable ø 5 ... 9 mm (0.20 ... 0.35 inches)
Technical data, special application cables
No. of conductors
Sqr. area
Shield
Color code
Outside color
Ext. diameter
Conductor
Isolation material
Flexible installation
Non-flexible installtaion
Capacity
Inductance
L/R
Metric measures (pressure in bar)
Flange rating
-5
PN 10
10.0
PN 16
16.0
PN 40
40.0
150 lb
19.7
Class D
10.3
8.3 Effect of temperature on working pressure
Coil cable
40 Ω
6 Ω
Coil cable
3
1.5 mm
2
Yes
Brown, blue, black
Grey
7.8 mm
Flexible CU
PVC
-5 ... +70°C
(23 ... 158°F)
-30 ... +70°C
(-22 ... 158°F)
161.50 pF/m
0.583 μH/m
43.83 þH/Ω
Temperature (°C)
10
50
10.0
9.7
16.0
15.5
40.0
38.7
19.7
19.3
10.3
10.3
Technical data
Standard electrode
cable
N/A
N/A
Special electrode cable
3
0.25 mm
2
Double
Brown, blue, black
Grey
8.1 mm
Flexible CU
PVC
-5 ... +70°C
(23 ... 158°F)
-30 ... +70°C
(-22 ... 158°F)
N/A
N/A
N/A
90
9.4
15.1
37.7
18.0
10.3
41

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sitrans f series

Table of Contents