Parker SciLog DINCon Installation, Operating,  & Maintenance Instructions
Parker SciLog DINCon Installation, Operating,  & Maintenance Instructions

Parker SciLog DINCon Installation, Operating, & Maintenance Instructions

Advertisement

®
®
®
SciLog
DINCon
, DINPres
,
®
& DINTemp
Modules
Installation, Operating & Maintenance Instructions

Advertisement

Table of Contents
loading

Summary of Contents for Parker SciLog DINCon

  • Page 1 ® ® ® SciLog DINCon , DINPres ® & DINTemp Modules Installation, Operating & Maintenance Instructions...
  • Page 2 Parker Hannifin Corporation. Parker Hannifin Corporation has a continuous policy of product development and although the Company reserves the right to change specification, it attempts to keep customers informed of any alterations. This publication is for general information only and customers are requested to contact your local Parker Sales Representative for detailed information and advice on a product’s...
  • Page 3: Table Of Contents

    Contents Introduction ..........................6 Calibration ..........................6 Warranty ............................. 7 Standards ........................... 7 Installation & Start-Up ......................... 7 Maintenance & Cleaning ......................7 DIN Module Specifications ......................8 Sensor Specifications ......................... 9 SciCon Sensors .......................... 9 SciPres Sensors ......................... 9 SciTemp Sensors ........................
  • Page 4 ALWAYS DISCONNECT this equipment from the power source before cleaning or performing maintenance. Contact your local Parker Division for parts, information and service. WARNING DISCONNECT ALL POWER TO THIS UNIT BEFORE INSTALLING, SERVICING, CLEANING OR REMOVING THE FUSE. FAILURE TO DO SO COULD RESULT IN BODILY HARM AND/OR PROPERTY DAMAGE.
  • Page 5 équipement. DÉBRANCHEZ TOUJOURS cet équipement de la source de courant avant de nettoyer ou d’exécuter l’entretien. APPELEZ PARKER pour pièces détachées, renseignements et entretien. ATTENTION DÉBRANCHEZ TOUT COURANT DE CETTE UNITÉ AVANT DE FAIRE L’INSTALLATION, D’EFFECTUER L’ENTRETIEN, LE NETTOYAGE OU AVANT DE RETIRER LE FUSIBLE.
  • Page 6: Introduction

    +1 805 604 3401 email: dhprocess@parker.com email: dhpsales.na@parker.com www.parker.com/dhsingleuse www.parker.com/dhsingleuse Parker customer service personnel will be able to serve you more efficiently if you have the following information:  Serial number and model name of the equipment  Installation procedure being used ...
  • Page 7: Warranty

    Warranty Country specific information can be found at: www.parker.com/termsandconditions Standards EN 61326-1:2006, Class B EN 61010-1:2010 Ed.3+ C1; C2 Conforms to UL STD 61010-1:2012 Ed.3+ R: 29Apr2016 Certified to CSA STD C22.2 No 61010-1-12:2012 Ed.3+U1;U2 Caution: The maximum recommended pressure for the sensors is 60 psi.
  • Page 8 Hardware Specifications – All Modules W: 0.875¨ (2.2 cm); Height 4.0¨ (10.2 cm); D: 4.75¨ (12.1 cm) Dimensions Weight 0.3 lb. (0.14 Kg) Molded Plastic Enclosure 24 VDC, 250 mA. Use only Class 2 External Power Supply Power CR1632, used for internal clock, not user serviceable Electrical Battery I/O Port Top...
  • Page 9 Specifications – Sensors SciCon: 5 VDC provided by DINCon ® Module Power Choice of five sizes: Luer, 3/8” Barb, 1/2” Barb, 3/4” Tri-Clover (TC) Sanitary and 1” Tri-Clover Fluid Connections (‘Ladish’) Sanitary. Polysulphone meets USP Class VI and FDA 21 CFR 177.1520, all wetted materials made with Wetted Materials animal-free compounds.
  • Page 10 +/- 0.30 psi at -5 to 30 psi for 2 point calibrated sensors (can be improved to -5 to 60 psi with 3 Accuracy point calibration, also available) 0.01 psi Resolution Temperature Range 0 to 60 °C (32 to 140 °F). EPROM, Stores Sensor ID, and Sensor Microchip Calibration Factor.
  • Page 11 Installation: Overview: The DINCon, DINPres, and DINTemp Modules are designed to be installed as part of a control system inside a control cabinet that has a 24 VDC source available and has Modbus communication protocol capability. The DIN Modules are installed on standard DIN rail. They are usually mounted with the red locking tab located at the bottom of the unit.
  • Page 12: Scicon Sensors

    All Sensors, SciCon for use with the DINCon Module, SciPres for use with the DINPres Module and SciTemp for use with the DINTemp Module are disposable flow through sensors that are purchased separately from Parker and come Pre-Calibrated. The Sensors connect to a provided cable that has a watertight connector on the sensor end and an 8 pin connector on the other end that plugs into a matching port on the bottom of the Module.
  • Page 13 DIN Rail locking clip Sensor 1 Sensor 2 Sensor 3 When powered up and connected the Sensor Status LEDs on the front of the Module are colored as follows:  Green = Connected  Red = Disconnected  Yellow = Connected but not calibrated. If the LED shows Yellow, it’s time to replace the affected Sensor with a new one.
  • Page 14 Operating Instructions: Modbus Overview: The Parker DIN modules communicate via the Modbus protocol. All data is transferred using the Modbus holding registers. The default connection is RS-485 at 115,200 baud, 8 bits, even parity and one stop bit. Note: If parity setting (register 109) is changed to “none” then two stop bits must be used.
  • Page 15 The Modules contain a Real Time Clock that will be set to the current date and the end users time zone prior to shipment from Parker. If re-setting the Real Time Clock is needed, the data must also be written in one continuous Modbus multi-register write operation to registers 130-137.
  • Page 16: Scipres Sensors

    DINTemp: Temperature Units: Default: ⁰C Register 140 T1 Temperature Limits: Low: Registers 160-161 High: Registers 162-163 T2 Temperature Limits: Low: Registers 164-165 High: Registers 166-167 T3 Temperature Limits: Low: Registers 168-169 High: Registers 170-171 Hardware Calibration: The factory hardware calibration data is available as read only values and is only performed on new modules with a special test fixture at the time of manufacture.
  • Page 17: Modbus Register List

    Modbus Register List: DINCon Register List: ADDRESS ACCESS FORMAT VALUE Default 90-99 ASCII String DINCon Serial Number 16 bit Model Number 0 = DINCon 101-102 IEEE Float Software Version 16 bit Reset Control bit 0 = System Reset bit 1 = Reset to Defaults bit 2 = Alarm Reset 104-105 16 bit...
  • Page 18 TCP Address P3 TCP Address P4 16 bit TCP Mask TCP Mask TCP Mask TCP Mask 16 bit TCP Gateway P1 TCP Gateway P2 TCP Gateway P3 TCP Gateway P4 16 bit RTCC Control bit0: Set Clock Values Registers 130-137 must be written as a block 16 bit Day of Week (1-7) 16 bit...
  • Page 19 bit3 = high alarm 154-155 IEEE Float Temperature Value 160-161 IEEE Float Conductivity Alarm Limit - Low 162-163 IEEE Float Conductivity Alarm Limit - High 164-165 IEEE Float Temperature Alarm Limit - Low 166-167 IEEE Float Temperature Alarm Limit - High IEEE Float Temperature Reference 25.0...
  • Page 20: Dinpres Registers

    DINPres Register List: ADDRESS ACCESS FORMAT VALUE DEFAULT 90-99 ASCII String DINPres Serial Number 16 bit Model Number 1 = DINPres 101-102 IEEE Float Software Version 16 bit Reset Control bit 0 = System Reset bit 1 = Reset to Defaults bit 2 = Alarm Reset 104-105 16 bit...
  • Page 21 16 bit TCP Mask P1 TCP Mask P2 TCP Mask P3 TCP Mask P4 16 bit TCP Gateway P1 TCP Gateway P2 TCP Gateway P3 TCP Gateway P4 16 bit RTCC Control bit0: Set Clock Values Registers 130-137 must be written as a block 16 bit Day of Week (1-7) 16 bit...
  • Page 22 bit 3 = High Alarm 151-152 IEEE Float P1 Pressure Value 16 bit P2 Status bit 0 = Disconnected bit 1 = Uncalibrated bit 2 = Low Alarm bit 3 = High Alarm 154-155 IEEE Float P2 Pressure Value 16 bit P3 Status bit 0 = Disconnected bit 1 = Uncalibrated...
  • Page 23 bit 2 = Start Span bit 6 = Cal in Progress bit 7 = Write Cal Data IEEE Float P1 Sensor Custom Pressure Offset (CPZ) IEEE Float P1 Sensor Custom Cal Factor (CCF) IEEE Float P1 Sensor Max Pressure 16 bit P1 Sensor MaxP, Day 16 bit P1 Sensor MaxP, Month...
  • Page 24 1 = 3/8" Barb 2 = 1/2" Barb 3 = 3/4" TC 4 = 1" TC 100 = Cal Fixture 16 bit P3 Sensor ID Material 0 = Polypropylene 1 = Polysulfone 16 bit P3 Sensor ID Month 16 bit P3 Sensor ID Year 16 bit P3 Sensor ID Lot...
  • Page 25: Dintemp Registers

    DINTemp Register List: ADDRESS ACCESS FORMAT VALUE DEFAULT 90-99 ASCII String DINTemp Serial Number 16 bit Model Number 2 = DINTemp 101-102 IEEE Float Software Version 16 bit Reset Control bit 0 = System Reset bit 1 = Reset to Defaults bit 2 = Alarm Reset 104-105 16 bit...
  • Page 26 16 bit TCP Mask P1 TCP Mask P2 TCP Mask P3 TCP Mask P4 16 bit TCP Gateway P1 TCP Gateway P2 TCP Gateway P3 TCP Gateway P4 16 bit RTCC Control bit 0 = Set Clock Values Registers 130-137 must be written as a block 16 bit Day of Week (1-7) 16 bit...
  • Page 27 bit 6 = Cal in Progress 157-158 IEEE Float T3 Temperature Value 160-161 IEEE Float T1 Temperature Alarm Limit - Low 162-163 IEEE Float T1 Temperature Alarm Limit - High 164-165 IEEE Float T2 Temperature Alarm Limit - Low 166-167 IEEE Float T2 Temperature Alarm Limit - High 168-169...
  • Page 28 16 bit T2 Sensor ID Unit Number IEEE Float T2 Sensor Cal Factor (CF) 16 bit T2 Sensor Custom Cal Status bit 0 = Invalid bit 1 = Start Zero bit 6 = Cal in Progress bit 7 = Write Cal Data IEEE Float T2 Sensor Custom Cal Factor(CCF) 16 bit...
  • Page 29 SciCon Disposable Conductivity / Temperature Sensors: CONNECTOR TYPE Max FLOW RATE Max PRESSURE NOMINAL CELL VOLUME Luer 1 liter/min 60 psi 80 µl 3/8" Barb 8 liters/min 60 psi 0.5 ml 1/2" Barb 17 liters/min 60 psi 0.8 ml 3/4" TC 31 liters/min 60 psi 1.2 ml...
  • Page 30 SciPres Disposable Pressure Sensors: CONNECTOR TYPE Max FLOW RATE * Max PRESSURE NOMINAL CELL VOLUME Luer 1 liter/min 60 psi 80 µl 3/8" Barb 8 liters/min 60 psi 0.5 ml 1/2" Barb 17 liters/min 60 psi 0.8 ml 3/4" TC 31 liters/min 60 psi 1.2 ml...
  • Page 31 SciTemp Disposable Temperature Sensors: CONNECTOR TYPE Max FLOW RATE * Max Pressure Luer 1 liter/min 60 psi 3/8" Barb 8 liters/min 60 psi 1/2" Barb 17 liters/min 60 psi 3/4" TC 31 liters/min 60 psi 1” Ladish TC 60 liters/min 60 psi *At a 1.0 psi drop across the sensor Specifications for all SciTemp™...
  • Page 32 SciCon Sensor Sanitization Test Protocol Parker/Scilog DN 3023: SciCon Sanitization, 1.0 Molar NaOH @ 22 ºC and 50 ºC, 2 Hours Group 1: SciCon Sensor Conductivity Response after Exposure to 1.0 N NaOH Solution, Temperature (1.0 N NaOH) at 22 º C, 2 Hours...
  • Page 33 Group SD* * SD = Standard Deviation NOTE: Parker sensors have been designed for disposable, single-use applications. However, with proper care, the sensors can be re-used repeatedly while maintaining good accuracy and precision. If required, sensors can be re- calibrated.
  • Page 34 SciPres Sensor Sanitization Test Protocol Parker/SciLog DN 3021: SciPres Pressure Sensor Sanitization Protocol (1): Sani- Sanitizing Conc. Temp. Pressure Contact Cycles Agent º C Time, Min 10 Cycles Limit NaOH 0.10 N 20-30 0 - 10 15-30 0.20% NaOH 0.50 N 20-30 13.5...
  • Page 35 For accurate performance, sensors should not be autoclaved more than two times. NOTE: Parker sensors have been designed for disposable, single-use applications. However, with proper care, the sensors can be re-used while maintaining good accuracy and precision. If required,...
  • Page 36 SciTemp™ Temperature Sensor Sanitization Test Protocols SciTemp Sanitization Protocol Sanitizing Conc. Temp. Pressure Contact Sani-Cycles Agent Limit º C Time, Min 0.10 N 20-30 0 - 10 15-30 NaOH NaOH 0.50 N 20-30 0 - 10 15-30 NaOCl 500 ppm 20-30 7 - 8 0 - 10...
  • Page 37 Summary: Sensor accuracy becomes increasingly compromised after the multiple autoclave cycles. For accurate performance, sensors should not be autoclaved more than two times. NOTE: Parker sensors have been designed for disposable, single-use applications. However, with proper care, the sensors can be re-used while maintaining good accuracy and precision. If required,...
  • Page 38 =1, none =2 3/26/10 A.Dawson Corrected error in RS485 A/B pin designation, it was reversed. Removed Bit 4 Error from Sensor Status Registers, it is unused. IOMI 02/07/18 A.Dawson Converted to Parker format, minor technical corrections.
  • Page 39 Notes...
  • Page 40 Notes...
  • Page 41 Notes...
  • Page 42 Parker Hannifin Corporation Parker Hannifin Manufacturing Ltd © 2015 Parker Hannifin Corporation. All rights reserved. Bioscience Division – N.A. Bioscience Division - EMEA Durham Road 2340 Eastman Avenue IOMI-AK1-004_RevB Birtley, Co. Durham Oxnard, California, USA 93030 DH3 2SF, England toll free: 877 784 2234...

This manual is also suitable for:

Scilog dinpresScilog dintemp