7. FUNCTION CHECK

8. TROUBLESHOOTING

Problem
Measurement does not
return plausible results.

9. SERVICING

 Perform functional check prior to each operation.
A complete functional check includes:
• Visual inspection for mechanical damage
• Visual and functional inspection of all movable pa-
parts
• Functional check of the control elements
CAUTION!
Risk of injury
If you notice any faults or deviations during the
functional check, do not use the unit.
 Have repairs of the unit carried out by seca
service or an authorized service partner.
Cause/Solution
Measurement result read off incorrectly
- Use the correct edge for the measuring range
currently in use ("Performing length measure-
ment" from page 14)
- Read off the decimal places for the current
measuring range in the correct direction ("Per-
forming length measurement" from page 14)
Incorrect position/posture of patient
- Correct the position/posture of the patient
("Performing length measurement" from
page 14)
Measuring rod mounted incorrectly on column
scale
- Mount the measuring rod correctly ("Before
you really get started..." from page 8)
The seca measuring rod does not require maintenance.
Should any part of it fail to function, please contact an
authorised service partner.
You can find your local service partner at
www.seca.com or just send an e-mail to
English
Function Check • 17