Adjusting position

To correct the position of the measuring rod proceed
as follows:
1. Slightly loosen the screw on the measuring rod.
2. Adjust the measuring rod together with the carriage
of the spacer according to the measured deviation
from the reference dimension.
3. Tighten the screw on the measuring rod.
4. Check the distance between the measuring flap
and patient standing area once again.
5. If necessary, re-adjust the position of the measuring
rod.
NOTE:
• If you are unable to set the specified reference
dimension, check that you have selected the
correct mounting position for your seca
column scale (see "Selecting the mounting
position" on page 9.).
• Your seca service partner will also be happy
to help. You can find your local seca service
partner at www.seca.com or just send an e-
mail to service@seca.com.