7. FUNCTION CHECK

8. SERVICING

► Perform a function check prior to each use.
A complete function check includes:
• visual inspection for mechanical damage
• checking the alignment of the device
• visual and function check of the display elements
• function check of all the controls shown in the section entitled
"Overview"
• function check of optional accessories
If you notice any faults or deviations during the function check, first try to
resolve the error with the aid of the section entitled "What do I do if ..." in this
document.
CAUTION!
Personal injury
If you notice any faults or deviations during the function check which
cannot be resolved with the aid of the section entitled "What do I do if
..." in this document, you may not use the device.
► Have the device repaired by seca service or by an authorized service
partner.
► Follow the section entitled "Servicing" in this document.
The bioimpedance analysis (BIA) measuring technology must be checked
every two years. We recommend servicing the whole device as part of this
check.
NOTICE!
Incorrect measurements as a result of poor servicing
► Have servicing and repairs carried out exclusively by seca service or
by an authorized service partner.
► You can find service partners in your area at www.seca.com or by
sending an e-mail to service@seca.com.
English
Function check • 43