Section 1 Specifications - Hach NO3D sc User Manual

Nitrate sensor
Table of Contents

Advertisement

Section 1
General Information
Measuring method
Measuring range
Minimum detection limit
Precision
Reproducibility
Response time (90%)
Measuring interval
pH range
Calibration methods
Power consumption
Power supply
Data transfer
Ambient data
Typical environment
Storage temperature
Operating temperature
Sample temperature
Max. flow velocity
Max. sensor immersion
depth/pressure
Max. air pressure for cleaning unit
option
General sensor information
Sensor dimensions
Sensor cable length
Sensor weight
Wetted materials
Installation angle
1
with standard solutions for ISE electrodes in lab conditions
2
®
Ryton
is a registered trademark of Phillips 66 Co.
Specifications
These are subject to change without notice.
Ion-selective electrodes for nitrate and chloride, pHD reference electrode and
temperature sensor
0.1 to 1000 mg/L [NO
0.5 mg/L [NO
5% of the measured value + 0.2 mg/L
5% of the measured value + 0.2
< 3 min (5 to 50 mg/L NO
continuous
pH 5 to pH 9
Sensor code for sensor cartridge
1 and 2-point matrix correction
1 W
via sc controller
via sc controller
Use in municipal waste water applications
Sensor: –20 to 60 °C (–4 to 140 °F)
Sensor cartridge: 5 to 40 °C (41 to 104 °F)
Air: –20 to 45 °C (–4 to 113 °F)
+2 to 40 °C (35 to 104 °F)
< 4 m/s
Can be immersed 0.3 to 3.0 m [1 to 10 ft]
max. pressure: 0.3 bar (4.4 psi).
3.1 bar (45 psi)
360.9 mm x 48.3 mm (14.21 x 1.9 in) (L x Ø), see
Standard: 10 m (33.8 ft)
Optional extension cables available in
5, 10, 15, 20, 30, 50 m [16.4, 33.8, 49.2, 65.6, 98.4, 164 ft].
Total maximum length: 100 m (328 ft)
Approx. 1326 g (46.77 oz)
Only for immersed installation: Sensor body made of stainless steel 316 with
ends made of Ryton PPS
45° +/– 15° vertical in flow direction
–N] and 0.1 to 1000 mg/L [Cl
3
1
–N]
3
1
mg/L1
-N)
3
®2
.
-
]
Figure 1 on page
6.
5