Hlai Models; Connect Wiring; Figure 3-5 Hlai Connection; Figure 3-6 Voltage Input Wiring - Honeywell OneWireless XYR 6000 Quick Start Manual

Hide thumbs Also See for OneWireless XYR 6000:
Table of Contents

Advertisement

3. Process Insertion

HLAI models

3.2
HLAI models

Connect wiring

Step
1
See Figure 3-5. Open the rear end cap (opposite end from display) by loosening the M3 locking
screw and unscrewing the end cap.
2
Open the cable gland (on right side below antenna). (Cable gland must be ordered as an option.)
3
Feed wiring (allowed diameter 6 to 8 mm) from other transmitter through the cable gland and
connect to terminal block using either voltage or current but not both. See Figure 3-6 or Figure
3-7. For hazardous locations see info starting on page 55.
4
Plug battery connector into batteries.
5
Close rear end cap and cable gland.
6
Tighten M3 locking screw on the rear end cap.
28

Figure 3-5 HLAI connection

4
4
1
1
5
5
2
2
6
6
3
3

Figure 3-6 Voltage input wiring

+
+
0-20 mA
0-20 mA
0-20 mA
4-20mA
4-20mA
4-20mA
-
-

Figure 3-7 Current input wiring

OneWireless XYR 6000 Transmitters Quick Start Guide