Handpump Assembly; Disassembly; Cleaning And Inspection; Re-Assembly - Fluke P3000 Series User Manual

Pneumatic deadweight tester
Hide thumbs Also See for P3000 Series:
Table of Contents

Advertisement

P3000 Series
Users Manual

Handpump Assembly

Disassembly

1. Disconnect tubing from check valve (19) and/or (23).
2. On the upper side of the top plate, unscrew locknut (3) 1/2 turn, and remove spring
3. The pump arm assembly can be swung out of the way.
4. Remove locknut (3) and clevis (2).
5. Loosen set screws (4) approximately 1 turn, and unscrew pump lock (5).
6. Withdraw pump assembly from below the top plate.
7. Disassemble outlet check valve assembly by unscrewing valve body (19). Take care
8. The inlet check-valve (23) is not user-serviceable and should be checked for signs of
9. Remove guide (6) from cylinder (13) — take care not to damage this, as it is a tight
10. Remove piston assembly from the cylinder.
11. The piston head assembly (12) can be separated from the piston shaft (10) by

Cleaning and Inspection

The piston head assembly cannot be further disassembled without damage to the
components, so it must be wiped clean with a soft cloth to remove any contamination.
Check all seals for signs of wear and/or damage.
Check the cylinder bore for signs of excessive wear, scratches/scoring, etc.
Replacement seals are available; however, if there is significant damage to the internal
components, the pump is available as a replacement assembly.

Re-Assembly

Reassembly is the reverse of the above disassembly instructions.
Examples:
5-10
info@Fluke-Direct.com
When handling the top plate assembly, it is good practice to
remove the piston assemblies to avoid accidental damage.
Replacement is simply the reverse of the above procedure.
clip (1) from clevis (2).
not to lose the small internal components that may spring out.
contamination, and replaced if necessary.
fit within the cylinder.
removing the snap ring (9).
Care should be taken whenever a seal comes in contact with
metal components, to avoid damage to the seal from sharp
edges, etc.
When introducing the piston head assembly (12) into the cylinder (13).
When introducing the piston shaft (10) through the guide (6).
When introducing the guide (6) into the cylinder (13).
Fluke-Direct
W Caution
Note
W Caution
.com
1.888.610.7664

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents