Page 1
Reference Manual 00809-0100-3550, Rev AA March 2019 ™ Rosemount 550pH Single-use pH Sensor for Bioprocessing Applications...
Page 2
The following instructions must be adhered to and integrated into your safety program when installing, using, and maintaining Emerson products. Failure to follow the proper instructions may cause any one of the following situations to occur: loss of life, personal injury, property damage, damage to this instrument, and warranty invalidation.
Reference Manual Sensor readiness 00809-0100-3550 March 2019 Sensor readiness Storage and handling Store the sensors attached to the single-use bioreactor until needed. WARNING Do not allow sensors to freeze or go below 39 °F (4° C). Disposal If the sensors have been contaminated, decontaminate prior to disposal. Please dispose with other electrical and electronic waste per local guidelines.
Page 6
Materials and units Electrical connector Variopol (VP 8.0) Sensitivity Greater than 57 mV / pH at 77 °F (25 °C) Electrolyte Phosphate based (see SDS on product website) Integral temperature sensor Pt-100 Gamma irradiation Up to 50 kGy Rosemount 550pH...
Reference Manual Install the sensor 00809-0100-3550 March 2019 Install the sensor The sensor will be integrated into a 1-in. barb fitting, sealed by several zip tie connectors. Note Complete Step 1 through Step 3 before filling the single-use bioreactor with solution. Procedure 1.
Page 10
4. Clip the zip tie, locking the clip in place. Remove the circular warning tag if it is still present. Figure 2-3. Figure 2-3: Clipped Zip Tie 5. Press the slide lock into the unlocked position. The small lip of the lock should be the side that is pressed in. See Figure 2-4. Rosemount 550pH...
Page 11
Reference Manual Install the sensor 00809-0100-3550 March 2019 Figure 2-4: Unlocked Slide Lock Note Before continuing to the next step, ensure the sensor is fully submerged in the solution. 6. Use the plunger assembly to push the sensor into the activated position. Figure 2-5.
Page 12
Figure 2-7: Zip Tie Postrequisites After the start-up is complete, there is no additional work needed on the sensor unless a one-point standardization is deemed necessary. See One-point standardization based on a grab sample for this procedure using an off-line sample. Rosemount 550pH...
1. See Figure 3-1 to wire the Rosemount 550pH Sensor to the Rosemount 56, 1056, or 1057 Transmitter. Figure 3-1: Wire the Rosemount 550pH Sensor to the Rosemount 56, 1056, or 1057 Transmitter 2. See Figure 3-2 to wire the Rosemount 550pH Sensor to the Rosemount 1066 Transmitter.
Page 14
Wire the sensor Reference Manual March 2019 00809-0100-3550 Figure 3-2: Wire the Rosemount 550pH Sensor to the Rosemount 1066 Transmitter 3. See Figure 3-3 to wire the Rosemount 550pH Sensor to the Rosemount 5081 Transmitter. Rosemount 550pH...
Page 15
Reference Manual Wire the sensor 00809-0100-3550 March 2019 Figure 3-3: Wire the Rosemount 550pH Sensor to the Rosemount 5081 Transmitter Reference Manual...
Page 16
Wire the sensor Reference Manual March 2019 00809-0100-3550 Rosemount 550pH...
Reference Manual Calibrate 00809-0100-3550 March 2019 Calibrate One-point standardization based on a grab sample Complete the following steps to perform a one-point calibration against an off-line ™ sample. This procedure is written for a Rosemount 56 transmitter. Procedure 1. Press ENTER on the transmitter. 2.
Page 18
Calibrate Reference Manual March 2019 00809-0100-3550 Rosemount 550pH...
Reference Manual Troubleshoot 00809-0100-3550 March 2019 Troubleshoot Faulty reading The one-point standardization at startup has a faulty reading. Recommended actions 1. Rotate sensor within sensor sleeve 90 degrees. 2. Repeat sensor standardization procedure. Label information too small The serial number, Cal pH, or slope are too small to read. Recommended action Scan the two-dimensional bar code on the sensor label to retrieve information.
Page 20
Troubleshoot Reference Manual March 2019 00809-0100-3550 Rosemount 550pH...
Reference Manual EU Declaration of Conformity 00809-0100-3550 March 2019 EU Declaration of Conformity Reference Manual...
Page 24
EU Declaration of Conformity Reference Manual March 2019 00809-0100-3550 Rosemount 550pH...
Page 25
Reference Manual 00809-0100-3550 March 2019 Reference Manual...
Page 26
2019 Emerson. All rights reserved. Facebook.com/Rosemount The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a youtube.com/RosemountMeasurement mark of one of the Emerson family of companies. All other marks are the property of their google.com/+RosemountMeasurement respective owners.