*RST Value
MAXimum
Examples
VOLT:HIGH 5
Query Syntax
[SOURce:]VOLTage:HIGH? [MINimum|MAXimum|DEFault ]
[SOURce:]VOLTage:LOW? [MINimum|MAXimum|DEFault ]
Returned Parameters
<NR3>
[SOURce:]VOLTage[:LEVel]:ON:HYSTeresis
This command sets the hysteresis value in the Living mode of the load.
Command Syntax
[SOURce:]VOLTage[:LEVel]:ON:HYSTeresis <NRf+>
Parameters
0 through MAX|MINimum|MAXimum|DEFault
Unit
V (volts)
*RST Value
MINimum
Examples
VOLT:ON:HYST 5
Query Syntax
[SOURce:]VOLTage[:LEVel]:ON:HYSTeresis? [MINimum|MAXimum|DEFault]
Returned Parameters
<NR3>
Related Commands
VOLT:LATCh
[SOURce:]RESistance[:LEVel][:IMMediate]
This command sets the resistance of the load when operating in constant
resistance mode.
Copyright © ITECH Electronics Co., Ltd.
Source Subsystem
61