Invacare ISA Series Service Manual page 9

Stand assist patient lift
Table of Contents

Advertisement

Model:
Checkpoints for Invacare® ISA™ according to ISO 10535:2006 Annex B
Functional tests
Castors smoothly swivel and roll
Application and release of castor brakes is working properly
Base fully opens and closes smoothly
Emergency stop engages when pressed in and all electrical functions are
disabled
Emergency stop can be reset when button is turned clockwise
Emergency functions on the control unit (emergency lowering and if
applicable emergency lifting) are working properly
Mechanical emergency release lowers the boom at safe speed
Successful load test (full lifting cycle with maximum safe working load
applied) with quite and smooth operation of the actuator
All buttons on the control unit and hand control are working and activate
the correct function
Battery indicator and LEDs on the control unit are working properly
Adjustment mechanism of the lifting arm is working properly (release,
smooth adjustment and self locking)
Adjustment mechanism of the leg support is working properly (release,
smooth adjustment and self locking)
Overall assessment
□ Pass
No safety or functional deficiencies were
detected.
Date of next inspection: _______________
Name:___________________________________
IMPORTANT
Corrective actions must be performed by a qualified technician.
Instructions for replacing and adjusting main parts are described in the service manual.
60121094-A
REF / SN number:
□ Fail
Safety or functional deficiencies detected and corrective action is
required.
Comments:
Date / Signature:____________________________________________
Comment
ü û
Servicing
2/2
9