KYORITSU KEW 6010B Instruction Manual page 10

Multi-function tester
Table of Contents

Advertisement

1.800.561.8187
● Operating instrumetal uncertainty of RCD (IEC 61557-6)
Function
×1/2
×1, FAST
Auto Ramp
The influencing variations used for calculating the operating error are denoted as follows:
Temperature : 0℃ and 35℃
Earth electrode Resistance (shall not exceed below) :
IΔn (mA)
10
30
100
300
500
System voltage : 230V+10%-15%
Supply voltage : 8V to 13.8V
Instrument dimensions:- 175 X 115 X 86mm
Instrument weight:-
Reference conditions
Battery type
Low battery warning
Operating temperature
and humidity.
Storage temperature
and humidity
Surge protection
Insulation resistance
www.
Operating uncertainty of trip current
-10% -
0%
0% -+10%
-10% -+10%
Earth electrode resistance (Ωmax.)
UL50V
2000
600
200
130
80
840g including batteries.
Specifications are based on the following conditions
except where otherwise stated:-
 1.Ambient temperature: 23± 5℃
 2.Relative humidity 45% to 75%
 3.Position: horizontal
 4.AC power source 230V, 50Hz
 5. DC power source: 12.0 V, ripple content 1% or
less
 6. Altitude up to 2000m, Indoor use
Eight R6 or LR6 batteries.
"
" symbol appears in the display if the battery
voltage drops below 8V.
0 to +40℃ , relative humidity 80% or less, no
condensation.
-20 to +60℃ , relative humidity 75% or less, no
condensation.
transient overvoltage 4000V
greater than 50MΩ at 1000V DC
(between enclosures and electrical circuits)
— 8 —
.com
UL25V
2000
600
200
65
40