Decontamination by autoclaving

NOTICE
IN
sartorius stedim
121°C
2 bar absolute
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. 25/11/2013 // 999281
The applicability of the decontamination method has to be ensured by the user.
Prior to decontamination clean surfaces with a damp cloth if necessary. Use water or
a mild detergent as cleaning agent.
Autoclaving
The collection bottle with bottle head and screw cap, the quick
coupling, the filter and the VHC
on hose are designated for steam sterilization at 121°C and
2 bar absolute (1 bar overpressure). Time of exposure accor-
ding to DIN 58946 t
Ensure not to exceed a temperature of 121°C.
Comply with instructions for use of the manufacturer of the
autoclave.
It is the user´s responsibility to ensure effective autoclaving.
Prior to autoclaving loosen or remove the bottle head
from the bottle.
Prior to autoclaving set the VHC
tion / autoclaving".
page 9 of 27
= 20 minutes.
e
NOTICE
NOTICE
with adapters and aspirati-
pro
to "continuous aspira-
pro