Decontamination of the pump parts in a disinfectant solution or by autoclaving

The applicability of the decontamination method and the selection of the ap-
NOTICE
propriate disinfectant have to be ensured by the user.
Prior to decontamination clean surfaces with a damp cloth if necessary. Use water or
a mild detergent as cleaning agent.
NOTICE
The vacuum pump has to be disassembled.
When disassembling avoid contact of contaminated parts with not contami-
nated parts.
NOTICE
The pictured pump parts have to be decontaminated.
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. 25/11/2013 // 999281
page 17 of 27
Tools required (metric):
- Diaphragm key SW 46 (order no.: 637777)
- Screw driver Torx TX20
- Open-ended wrench SW 19