Function Reference
Channel Quality Measurement Function
Error Code
hpe5022_ERROR_EXCESS_TRACK_DATA
hpe5022_ERROR_HARD_HAMP
hpe5022_ERROR_INV_ID
hpe5022_ERROR_INV_DRIVE_CONDITION
hpe5022_ERROR_INV_PARAMETER
hpe5022_ERROR_NOT_INIT
See Also
"hpe5022_BER_channelQuality_Q" on page 165
212
Description
The data overflows a track at one
revolution. Change the setting of the
user data rate, rpm , track number,
track format or sector format.
Hardware error is detected in the
head amplifier.
The handle specified by 'id' is
invalid.
The spinstand drive is turned off.
Turn it on before executing this
function. See the
"hpe5022_driveState" function.
The parameters 'seqType' is out of
range.
The resource of Agilent E5039A/B
can not be opened during
initialization. Check if Agilent
E5039A/B is included in the
rsrcArray of the "hpe5022_init"
function.
Chapter 3