Reconditioning; Cleaning And Disinfection; Material Overview; Reconditioning Scheme - Invacare 9630E Styxo Service Manual

Hygiene medical devices
Hide thumbs Also See for 9630E Styxo:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 3

5 Reconditioning

5.1 Cleaning and disinfection

IMPORTANT!
All cleaning agents and disinfectants used must
be effective, compatible with one another
and must protect the materials they are
used to clean. For further information on
decontamination in Healthcare Environments,
please refer to your local guidelines on infection
control and your local infection control policy.
Preparation at the place of use
1. Thorough pre-clean the product with hot tap water.
– Pre-cleaning storage time must be as short
as possible.
– Products to be cleaned must be protectively
packed, e.g. sealed with adhesive tape in
such a way that no more openings are visible
on the packaging.
– Pathogens with which the product has been
infected (even potentially) must be noted and
visibly specified on the outer packaging.
Preparation before cleaning
1. Dismantle the product into individual components and
create an optimum item for cleaning and disinfection.
2. If present, cover electrical connections with protective
caps and seal any sites where water may leak in.
3. Remove superficial contamination.
Cleaning
1. Clean the product with decalcifying sanitizing
detergent. Manual brushing is permitted.
2. Rinse with tap water until the detergent is completely
removed.
3. Mechanically cleaning of washable components (e.g.
covers) in industrial washing machine.

5.3 Reconditioning scheme

For replacement parts refer to the electronic spare parts catalogue (ESPC).
The following items must be reviewed and checked when reconditioning is required:
Toilet frames
Issue
Products
Styxo 9630E
Frame damaged (cracks,
fractures)
Frame coating damaged or
scratched.
Toilet seat damaged
(cracks, sharp edges)
Toilet pan and/or lid
damaged
Splash guard damaged
Armrest pad damaged
1641383-A
4. Rinse with tap water until the detergent is completely
removed.
5. Manually clean the product with universal detergent.
Supplementary manual brushing is also permitted.
6. Rinse with tap water until the detergent is completely
removed.
7. Air dry until there is no more superficial moisture on
the product.
Disinfection
1. Disinfect the product by wiping generally accessible
surfaces with disinfectant.
2. Air dry until there is no more superficial moisture on
the product.

5.2 Material overview

Products
Styxo
Aquatec 90
Aquatec
900
Aquatec
90000
Pico / Pico
TS
Tellus /
Pluto
Sorrento
range
Troubleshooting action
Discard the product.
Use touch-up colours to repair damages. See chapter 4.3 Colour
codes, page 10.
Replace with new one. See ESPC.
Replace with new one. See ESPC.
WARNING!
Risk of contamination
– Do not continue to use the product in
case of water penetration or if it shows
discolourations which cannot be removed.
Material(s)
PP, Coated steel
PP
PP, Coated steel
PP
PP, Aluminum
PP, Coated steel
PP (glass fiber reinforced)/TPE, Aluminum,
Nylon (PVC coated)
Reconditioning
OK
11

Advertisement

Table of Contents
loading

Table of Contents