6 General repair information

IMPORTANT!
– If possible, continue to use the old
identification label; if this is not possible,
the new identification label must contain
the same information and the old serial
number. (Replacement of spare parts with
serial numbers).
– When components are replaced it is necessary
to ensure the traceability of the components
replaced.
– If screws with thread locker are loosened,
these must be replaced with new screws
with thread locker. Alternatively, new thread
locker must be applied (e.g. Loctite 2701).
– Parts that become damaged during removal
must be replaced with new ones.
14
For instructions on how to replace the spare parts refer
to the graphics in the ESPC.