Dickson PR350 Manual

Dickson PR350 Manual

Small pressure data logger

Advertisement

Phone: 1-800-757-3747 or 1-630-543-3747 | Fax: 1-630-543-0498 | Email: support@dicksondata.com | Website:DicksonDate.com

Getting Started

1. Install version 17.1.6 or higher of DicksonWare Software
2. Unscrew the logger cover; connect the cable (supplied with the software) to the logger and to a working USB port on your
computer.
3. Click the Setup button. The Setup window will appear, and all fields should be automatically filled in, this will confirm that
DicksonWare has recognized the logger. Should all fields remain blank, refer to "No Communication" in the Support section
at www.dicksondata.com for this logger model. Once DicksonWare recognizes the logger, press the CLEAR button. This will
delete all data currently stored.
4. The logger is now sampling and ready for use. The default sample interval is 5 minutes. You may choose to change the
default setting for sample interval and start date and time. When changing these settings. The logger will automatically go
through the CLEAR process to accept your changes. NOTE: To conserve battery life when storing the unit, press the
"Setup" button and select "Stop When Full"
WARNING: Fire, explosion, and severe burn hazard. Do not recharge, disassemble, heat above 85°C, or incinerate or expose
internal battery to water

Useful Features

1. Delayed Start: Program your logger to the date/tim Delayed Start: e to begin logging.
2. Real Time Monitoring: Monitor and graph your proce Real Time Monitoring: ss as it happens and print and save from real
time screen.
3. Customize Graphed Data: DicksonWare calculates Min Customize Graphed Data: , Max and Average of all data collected.
Customize data by highlighting specific data points which will recalculate MIN, MAX and Average for highlighted information.
4. Exported Data: A snapshot of your graph or a real t Exported Data: ime graph of points can be easily exported to other
programs such as Excel or PowerPoint.
Click on the FEATURES button in the software for detailed instructions on how to use above features.
Small Pressure Data Logger
PR150, PR350

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PR350 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Dickson PR350

  • Page 1 PR150, PR350 Phone: 1-800-757-3747 or 1-630-543-3747 | Fax: 1-630-543-0498 | Email: support@dicksondata.com | Website:DicksonDate.com Small Pressure Data Logger Getting Started 1. Install version 17.1.6 or higher of DicksonWare Software 2. Unscrew the logger cover; connect the cable (supplied with the software) to the logger and to a working USB port on your computer.
  • Page 2 Level Monitor on the Setup screen displays the battery voltage and a Low Battery Warning when replacement is needed. The battery is user replaceable and MUST be replaced with one of the same type. A Tadiran TLH-5955, XENO XL-055F or equivalent is required. Dickson sells replacement batteries. IMPORTANT: Be sure to observe correct polarity when installing the new battery.

This manual is also suitable for:

Pr150

Table of Contents