Welch Allyn KleenSpec 800 Series Directions For Use Manual page 20

Cordless illumination system
Hide thumbs Also See for KleenSpec 800 Series:
Table of Contents

Advertisement

16 Appendices
Recommended separation distances between portable and mobile RF communications
equipment and the 800 Series
The 800 Series is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or user of the 800 Series can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the 800 Series as recommended
below, according to the maximum output power of the communications equipment.
Rated max. output
power of transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distanced in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
Test specifications for enclosure port immunity to RF wireless communications equipment
Test frequency
(MHz)
385
450
710
745
780
810
870
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
150 kHz to 80 MHz in
outside ISM bands
ISM bands
0.12
0.20
0.37
0.63
1.17
2.00
3.69
6.32
11.67
20.00
a
a
Band
Service
MHz
380 - 390
TETRA 400
430 - 470
GMRS 460,
FRS 460
704 - 787
LTE band 13, 17 Pulse modulation
800 - 960
GSM 800/900,
TETRA 800,
®
Welch Allyn
KleenSpec 800 Series Cordless Illumination System
80 MHz to 800 MHz
0.12
0.38
1.20
3.79
12.00
Maximum power
b
Modulation
(W)
1.8
b
Pulse modulation
18 Hz
c
2
FM
±5 kHz
deviation
1 kHz sine
b
0.2
217 Hz
b
2
Pulse modulation
18 Hz
800 MHz to 2.7 GHz
0.23
0.73
2.30
7.27
23.00
Distance (m)
Immunity
test level (V/
m)
0.3
27
0.3
28
0.3
9
0.3
28