Scion Instruments 436-GC Service Manual page 69

Gas chromatograph
Table of Contents

Advertisement

436-GC/456-GC
Code
Error Description
428
Sampler Tower stuck on sensor
429
Sampler Carrousel stuck on sensor
434
Sampler EDS Checksum is bad
436
IOP Communication failure
437
Battery voltage out of range
439
PGA Calib Failed
440
Temp Calib Failed
441
Temperature Calibration Failed
442
Press out of tolerance
443
Boardtemp out of tolerance
Out of Tolerance (red_id)
444
P10V ref out of tolerance
445
AIN10V out of tolerance
446
P5V ref out of tolerance
447
P2V5 ref out of tolerance
448
P24V Prim out of tolerance
449
P24V Sec out of tolerance
450
N24V out of tolerance
451
P5V D out of tolerance
452
P5V out of tolerance
453
N5V out of tolerance
454
P15V out of tolerance
455
N15V out of tolerance
456
P5V DAC out of tolerance
457
VREF1 out of tolerance
458
VREF0 out of tolerance
Out of Tolerance (cts_id)
459
CTS1 out of tolerance
460
CTS2 out of tolerance
461
CTS3 out of tolerance
462
CTS4 out of tolerance
463
CTS5 out of tolerance
464
CTS6 out of tolerance
465
Tref out of tolerance
466
Step motor moving
Temperature
470
Maximum Setup Temperature Exceeded
471
Maximum Temperature Exceeded
473
Minimum Temperature Exceeded
474
Open Sensor
476
Shorted Sensor
477
Thermal Runaway
478
Too Fast Response
479
Too Slow Response
SCION Instruments
436-GC/456-GC Service Manual Revision B February 2019
Error level
Cause Condition(s)
Alert
(errors not reproducible)
Alert
(errors not reproducible)
Alert
-
Error
IOP Communication failure
Battery voltage < 2V or > 4V  error
Event
Error
IOPSH (measuring system)
Error
-
Ambient pressure < 15kPa or > 115pKa  error
Error
Error
Currently in use, only logged
444 to 465 Refvolt out of Tolerance (all voltages on
Error
mainboard, part of measuring system) Hazardous by
default, basis of all measurement in system.
Error
Error
Error
Error
Error
Error
Error
Error
Error
Error
Error
Error
Error
Error
Error
Error
Error
Error
Error
Error
Error
Error
GC action
Electronics off (all)
H2 flow set to 0.0 mL/min
Electronics off (all)
H2 flow set to 0.0 mL/min
-
Notification Message?
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
Electronics off (all)
-
H2 flow set to 0.0 mL/min
-
-
-
-
-
-
-
Page: 68
510 Error codes
Action needed
-
-
-
Replace/fix IOP
-
Recalibrate PGA
Recalibrate temp
-
GC main board
Check power supply requirements/
replace main board electronics
-
-
-
-
-
-
-
-

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for Scion Instruments 436-GC

This manual is also suitable for:

456-gc

Table of Contents