Fan-Motor Adjustment

This procedure must be carried out after replacement and/or removal of the Fan-Motor.
If not carried out correctly, the Fan-Motor can be damaged in time.
1. Turn all 4 inner nuts hand tight against the motor housing.
2. Place the Fan-Motor on the motor mount.
3. Turn the 4 inner nuts against the motor mount (hand tight!).
4. Check and correct if necessary (with the inner nuts) if the fan motor housing is
parallel with the motor mount.
SCION Instruments
436-GC/456-GC Service Manual Revision B February 2019
410 Procedure
Page: 48