436-GC/456-GC
Code
Error Description
142
Aux 3 EFC Hardware Error
143
Aux 4 EFC Hardware Error
144
Aux 5 EFC Hardware Error
145
Aux 6 EFC Hardware Error
146
Aux 7 EFC Hardware Error
147
Aux 8 EFC Hardware Error
TCD no Air Leak (position)
230
Front TCD no Air leak
231
Middle TCD no Air leak
232
Rear TCD no Air leak
TCD Air Leak fault (position)
233
Front TCD Air leak fault
234
Middle TCD Air leak fault
235
Rear TCD Air leak fault
Max Setup Temper Exceeded (position)
247
Column Oven Max Setup Temper Exceeded
248
Zone 1 Max Setup temper Exceeded
249
Zone 2 Max Setup temper Exceeded
250
Zone 3 Max Setup temper Exceeded
251
Zone 4 Max Setup temper Exceeded
252
Zone 5 Max Setup temper Exceeded
253
Zone 6 Max Setup temper Exceeded
Temperature Zone
260
Column Temperature Out Of Tolerance
261
Zone 1 Temperature Out Of Tolerance
262
Zone 2 Temperature Out Of Tolerance
263
Zone 3 Temperature Out Of Tolerance
264
Zone 4 Temperature Out Of Tolerance
265
Zone 5 Temperature Out Of Tolerance
266
Zone 6 Temperature Out Of Tolerance
Detector EFC Flow out of range
304
Front DFC Chan1 out of range
305
Front DFC Chan2 out of range
If the used gas is Hydrogen (H
) the Error level is Error, in all other cases an Alert is given.
2
2
SCION Instruments
Error level
Event
Event
Event
Alert
Alert
Alert
Error
Error
Error
Error
Error
Error
Error
Alert Error 2
Alert/ Error
436-GC/456-GC Service Manual Revision B February 2019
Cause Condition(s)
When error (18) is solved within 4 minutes after error
(18) appears
When error (18) is solved within 4 minutes after error
(18) appears
When error (18) is solved within 4 minutes after error
(18) appears
4 minutes after error (18) where there has been no
"No Air Leak"
4 minutes after error (18) where there has been no
"No Air Leak"
4 minutes after error (18) where there has been no
"No Air Leak"
Defect Sensor. Temperature is
10 °C above set temperature limit
Defect Sensor. Temperature is
10 °C above set temperature limit
Defect Sensor. Temperature is
10 °C above set temperature limit
Defect Sensor. Temperature is
10 °C above set temperature limit
Defect Sensor. Temperature is
10 °C above set temperature limit
Defect Sensor. Temperature is
10 °C above set temperature limit
Defect Sensor. Temperature is
10 °C above set temperature limit
Flow is above max +10 % or below -5 mL/min for
more then 10 sec.
Flow is above max +10 % or below -5 mL/min for
Page: 65
GC action
Action needed
Notification Message
-
Notification Message
-
Notification Message
-
Turn Electronics and power
Find source and repair/replace
off
hardware
Turn Electronics and power
Find source and repair/replace
off
hardware
Turn Electronics and power
Find source and repair/replace
off
hardware
Electronics off (all)
Check Column Oven zone
H2 flow set to 0.0 mL/min
Electronics off (all)
Check zone (1)
H2 flow set to 0.0 mL/min
Electronics off (all)
Check zone (2)
H2 flow set to 0.0 mL/min
Electronics off (all)
Check zone (3)
H2 flow set to 0.0 mL/min
Electronics off (all)
Check zone (4)
H2 flow set to 0.0 mL/min
Electronics off (all)
Check zone (5)
H2 flow set to 0.0 mL/min
Electronics off (all)
Check zone (6)
H2 flow set to 0.0 mL/min
Turn off/on flow, check source
Activate Safety Method
(gas-supply)
Activate Safety Method
Turn off/on flow, check source
510 Error codes