Useful Functions (Calibration Offset Function) - Yamato IC103C Manual

Economy incubator
Hide thumbs Also See for IC103C:
Table of Contents

Advertisement

Using the calibration offset
function
MEASURED TEMP.
HEATER
RUN
SET TEMP.
TIMER
SUB MENU
Calibration offset function compensates any differences
between the target temperature in the chamber and the
control temperature of the controller (sensor temperature.)
The function can compensate in parallel to either plus or
minus side for the whole temperature band of the unit.
The lock can be set or released with the SUB MENU keys.
The temperature is set at "0" on shipping from the
factory.
① Start operation at the target set temperature and confirm
the temperature in the chamber with a temperature
recorder after temperature has stabilized.
② Confirm the difference between the set temperature and
that in the chamber.
③ Press the TIMER key (SUB MENU key) long to enter the
RUN
sub menu mode.
STOP
Press the TIMER key (SUB MENU key) several times to
select the characters cAL
calibration offset function.
④ Enter the difference between the set temperature and
the temperature in the chamber using the ▼▲ keys
and press the TIMER key (SUB MENU key) long to exit
the sub menu mode. (When you want to set the key
lock function, proceed to character selection process for
the key lock function without pressing the TIMER key
(SUB MENU key) long. )
4. Operating procedures

Useful functions (calibration offset function)

26
Control
temperature after
minus side
compensation
Current
temperature
Control
temperature after
plus side
compensation
that indicates the