Sensor Specifications

The sensors used in this instrument are listed below and operate with the
measurement ranges, resolutions, accuracies, temperature ranges, and humidity
ranges listed. The accuracies stated below are over the entire operating range of
the sensor and defined over the range of calibration.
Sensor Specifications for the GasBadge Pro
Range
CO
0 – 1,500 ppm
CO/H2
(Hydrogen
0-1000ppm
Low)
H
S
0 – 500 ppm
2
O
0 – 30% vol
2
NO
0 – 150 ppm
2
SO
0 – 150 ppm
2
NH
0 – 500 ppm
3
Cl
0 – 100 ppm
2
ClO
0 – 1 ppm
2
PH
0 – 10 ppm
3
% Accuracy
At
temperature
of calibration
(measure-
ment
subrange)
± 5.0
1 ppm
1ppm
± 5.0
±5% (0.0 to
200.0 ppm)
0.1 ppm
±15% (200.1 to
500.0 ppm)
± 0.5
5
0.1%
± 10.0
0.1 ppm
±8% (0.0 to
10.0 ppm)
0.1 ppm
+8% to +20%
(11.0 to 150.0
ppm)
± 15.0
1 ppm
±10% or 0.2
6
ppm
(0.0 to
10.0 ppm)
0.1 ppm
± 15.0 (11 to 50
ppm)
±10% or 0.05
0.01 ppm
6
ppm
±6% or 0.1
0.01 ppm
6
ppm
72
1
Over full
measurement
temperature
3
ranges
2
± 15
± 15
± 15
± 0.8
5
± 15
± 15
± 15
± 15
± 15
± 15
Sensor
Temp
% RH
4
Range
Range
°C (°F)
-40 to +50
15 – 90 %
(-40 to 122)
-20 to +50
15 – 90 %
(-4 to 122)
-40 to +50
15 – 90 %
(-40 to 122)
-20 to +50
0 – 99 %
(-4 to 122)
-20 to +50
15 – 90 %
(-4 to 122)
-40 to +50
15 – 90 %
(-40 to 122)
-20 to +40
15 – 90 %
(-4 to 104)
-20 to +40
10 – 95 %
(-4 to 104)
-20 to +40
15 – 90 %
(-4 to 104)
-20 to +40
20 – 95 %
(-4 to 104)