Table 1:
Immunity (all devices): electromagnetic emissions
Guidance and manufacturer's declaration – electromagnetic immunity
The ergo vac is intended for use in the electromagnetic environment specified below.
The customer or the user of s ergo vac hould assure that it is used in such an environment.
Immunity test
standard
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions and
voltage variations on power supply
lines
IEC 61000-4-11
Note: U
is the a.c. mains voltage prior to application of the test level.
T
Power frequency (50 / 60 Hz)
magnetic field
IEC 61000-4-8
IEC 60601
Compliance
test level
level
± 8 kV contact
± 8 kV contact
±15 kV air
±15 kV air
± 2 kV for power
supply lines
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 1 kV line to line
± 1 kV line to line
± 2 kV line to earth
± 2 kV line to earth
<5% U
(0,5 cycle)
<5% U
T
40% U
(5 cycles)
40% U
T
70% U
(25 cycles)
70% U
T
<5% U
for 5 s
<5% U
T
3 A / m
200 A / m
User guide
Electromagnetic
environment - guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
(0,5 cycle)
T
(5 cycles)
Mains power quality should be that of a
T
typical commercial or hospital environ-
(25 cycles)
ment.
T
for 5 s
T
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hos-
pital environment.
ergo vac
Seite 37