7.2.2
Recommended Cleaning and Disinfection Methods for Electrodes
This cleaning method can damage the suction unit if not carried out correctly.
Caution
Instructions for Wiping Disinfection / Cleaning
1. Use only the disinfectants mentioned in Point 5.3.1.
2. Clean / disinfect all areas of the electrode, which have come into contact with the pa-
Electrode suction dome out-
side, grip area / suction area
Suction area
Diagram 1:
Suction dome cleaning
outside
Attention: if the inside is cleaned incorrectly, particles (skin flakes, contact product
residues) can remain in the area of the sealing lip (see Diagram 3)
Wiping disinfection / cleaning: to be carried out after every use
Intensive wiping disinfection / cleaning: 1x daily after the last use
Immersion disinfection, cleaning / drying: 1x daily after the last use
tient.
Cleaning of the
electrode contact surface
Diagram 2:
Clean contact surfaces
User guide
Electrode suction dome inside,
sealing lip, electrode body /
suction dome, suction area
Sealing lip
Suction area
Diagram 3:
Suction dome cleaning inside
ergo vac
Seite 23