Setting the Device Parameters

(DEVICE)
Purpose
Device configuration, i.e., to meet
individual requirements by selecting
predefined menu parameters in the
Setup menu. You can block access to
the menu by assigning a password.
Features
The device parameters are combined in
the following groups:
– Password to the Setup menu
– Interface
– Display contrast
– Acoustic signal
– Factory settings
You can display, enter or change the
following parameters:
Password
– Password for access to the device
parameters and isoTEST functions
in the Setup menu, and to drying
programs in the Program menu
Interface
– SBI operating mode
Simple record of analysis results for PC
or external printer, factory setting for
the YDP03-0CE printer
Format: baud rate, number of data bits,
parity, stop bits, handshake
– xBPI operating mode
Function-oriented interface with trans-
parent data transmission
Network address: enter a number from
0 to 31; factory setting: 0
Display contrast
– Contrast/angle of the display (enter a
number from 0 to 4; factory setting: 2)
Acoustic signal
– Acoustic signal on or off
14
Factory Settings
Parameters: The factory-set configura-
tions are identified by an "ο" in the list
starting on page 17.
Preparation
Display existing device parameters
§ Select the Setup menu:
Use the x or y key to select
SETUP in the Function line and press
u to confirm
> SETUP is displayed:
§ Select Device parameters: Press the
y key and press u to confirm
If no password has been assigned,
anyone can access "SETUP: Device
parameters" without entering a
password
If a password has been assigned:
> The password prompt is displayed
$ If the Setup menu is protected by a
password: Use the x and y keys
to enter the password (see page 9 for
details) and press u to confirm
§ Confirm the password and display
"Device parameters":
Press the u key
> Device parameters are displayed:

Entering or Changing the Password

– The password (up to 8 characters)
protects the following functions:
– Setup: Device parameters
– Program: Drying programs
– Setup: isoTEST functions
§ Display Device parameters (see
"Preparation > Display existing device
parameters" above)
§ Write down the password here for easy
reference:
Password = . . . . . . . . . . . . . . . . . . . . . .
If you assign a password and then
forget what it is:
$ Enter the General Password
(see Appendix)
$ Confirm the password and display
device parameters:
Press the u key
> Parameters are displayed (see above)
§ Configuring device parameters:
Select "Password" (if configured) and
confirm: Press the u key
> Password: and any existing password
are displayed:
§ New password: Use the x and y
keys to enter the new password (up to
8 characters) and press u to confirm
If no entry is shown under "Password",
then no password has been stored
§ To confirm your input:
Press the u key
§ Exit the Setup menu:
Press the c key twice