Accuracy Specifications
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, relative humidity at 0 % to
75 %. Accuracy specifications take the form of: ±([% of Reading] + [Number of Least Significant Digits]).
AC and DC Voltage
Function
AC Volts (40 Hz – 500 Hz)
AC Millivolts
DC Millivolts
DC Volts
[1] All ac, Hz, and duty cycle are specified from 1 % to 100 % of range. Inputs below 1 % of range are not specified.
Range
4.000 V
[1]
40.00 V
400.0 V
1000 V
400.0 mV
400.0 mV
4.000 V
40.00 V
400.0 V
1000 V
Resolution
15B+
0.001 V
0.01 V
1.0 % + 3
0.1 V
1 V
0.1 mV
3.0 % + 3
0.1 mV
1.0 % + 10
0.001 V
0.01 V
0.5 % + 3
0.1 V
1 V
Digital Multimeters
Accuracy Specifications
Accuracy
17B+
18B+
1.0 % + 3
1.0 % + 3
3.0 % + 3
3.0 % + 3
1.0 % + 10
1.0 % + 10
0.5 % + 3
0.5 % + 3
19