General Information; Continuous Operation – Mode Of Operation; Electrical Rating; Power Table Specifications - Zeiss Stratus OCT 3000 Service Manual

Table of Contents

Advertisement

Stratus OCT Model 3000 Service Manual

General Information

`ä~ëëáÑáÅ~íáçå=L=`çãéäá~åÅÉ=
Class I Equipment – Protection against electrical shock.
Class I Equipment
Class I Equipment
Class I Equipment
Type B – Degree of protection against electric shock of applied
Type B
Type B
Type B
part (chin and forehead rests).
Ordinary Equipment (IPXO) – Degree of protection against ingress of liquids (none).
CAUTION: This instrument has no special measures to protect
CAUTION:
CAUTION:
CAUTION:
against harmful ingress of water or other liquids (classified IPXO –
ordinary equipment).
Do not place containers of liquid on or near the instrument, and
do not use aerosols on or near it.
Continuous Operation – Mode of operation.
Not suitable in the presence of flammable anesthetic.
The Stratus OCT Model 3000 complies with the EC Directive 93/42/EEC on
Medical Devices however it does not comply with the EC Directive 2011/65/EU
on RoHS (Recast) therefore it cannot be CE-Marked.
The Stratus instrument / review software is compliant with the EC Directive
93/42/EEC on Medical Devices and is not required to be compliant with EC
Directive 2011/65/EU on RoHS (Recast) therefore it will be CE-Marked.
=
Complies with US and Canadian medical electrical system safety requirements.
=
=
póëíÉã=bäÉÅíêáÅ~ä=fåÑçêã~íáçå=
=

Electrical Rating:

100 – 120V
~
220 – 240V
~

Power Table Specifications

Electrical rating: 120 V~, 60 Hz, 2.5A, 300 W
Loading: 1000 N Push
Duty Cycle: Int. 0.8 min / 21 min
Protection against Ingress: IP20
2
, 50/60 Hz, 5A Max
, 50/60 Hz, 2.5A Max
2660021155438 A

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents