Menu: Number of samples, press the
The display will read "Number of samples xxx". Replace xxx by a number from the range of 1 to 255 (usually
25 is chosen). Move the cursor using the
respectively. Confirm the selection by pressing the
any key". Press the
key or any other key. This action will take you back to the basic menu.

7.1.3. Analog output

Setting options
Connected to terminals 4 and 5 is a programmable current output. It is an active current output, galvanically
isolated from other meter parts. The maximum output load is 1,000Ω. Depending on the liquid-flow
characteristics, the output can be used in four different modes of operation (see graphics below) and in two
selectable measurement ranges.
When the flowmeter is used for measurement of flow in both directions (bi-directional measurement), we
recommend to use current output mode |Q| and use one of the binary outputs (OUT1, OUT2 or relay output)
for indicating the flow direction.
Please contact the producer if is necessary to use mode –Q .. +Q and get information about the flow rate
and direction of flow at the same time (when using –Q .. +Q mode, there is no error indication for exceeding
analog range in negative flow direction).
The two-line display unit will always show two of the items from the following menu. Use the
to browse downwards and upwards through the menu items.
Menu: Analog output, press the
0...+Q Output
0...-Q Output
|Q| Output
–Q...+Q Output
Fixed current 0...20
In all operating modes except the "Fixed Current" mode, the current output range can be user defined.
Menu: Analog output, press the
Output 0...20mA
Output 4...20mA
Selection of current output
Menu: Analog output, press the
Flow rate for Imax
The current output setting consists of defining flow rate Q
key, and increase/decrease the Q
setting by pressing the
preferable the
key. This action will take you back to the main configuration menu, item "Analog output".
In the "Fixed current" mode, the output current can be set within the range of 0 to 20 mA. This mode is used
for meter servicing purposes.
Menu: Analog output, press the
Move the cursor using the
the desired current value and confirm the setting by pressing the
Entered Press Any Key". Press any key, preferably the
menu, item "Analog Output". At the same time, the defined current will start to flow through the output circuit.
ELIS PLZEŇ a. s., Luční 425/15, 301 00 Plzeň, Czech Republic, Phone: +420/377 517 711, Fax: +420/377 517 722
Design, Assembly and Service Manual
Electromagnetic flowmeter FLONET FN20xx.1
Suppression of step changes in flow rate
key.
key and increase/decrease this number using the
key
key / Output 0...+Q, press the
key / Output 0...+Q, press the
value using the
4
key. The display will then read "Value Entered Press Any Key". Press any key,
key / Fixed current 0...20, press the
key, and increase/decrease the current value using the
key. The display will then read "Value entered Press
key
key / Output 0...20mA, press the
corresponding to Imax. Move the cursor using the
4
and
keys. Select the desired Q
key / Fixed current 0...20 mA
key. The display will then read "Value
key. This action will take you back to the main
Page 30 of 69
or
keys ,
and
keys
key /
and confirm the
4
and
keys. Select
Es90420K/c

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents